Product Overview
Rugged Shock Transportation Data Logger with 2 acceleration sensors ±15 g & ±200 g, air cargo approved, stores up to 2 mill. readings
Features Include:
Whether it be to verify questions of liability, detect damage early or to optimise packaging, the MSR175 data logger accurately records critical transport events such as shocks, jolts or fluctuations in temperature, pressure or humidity and provides clarity with objective measured data.
The two digital 3-axis-acceleration sensors (± 15 g and ± 200 g) integrated in the MSR175 record shocks at a measurement frequency of up to 6400/s. The storage capacity of over 2 million measured values is sufficient to record at least 1,000 shocks (230 mAh Li-Po battery) or approximately 1,000 shocks (3.6 V, 7700 mAh Li-SOCI2 battery), depending on the type variant. In addition, the transportation data logger measures and records temperature profiles from -20° to +65 °C. Further type variants of the MSR175 are available with internal humidity, air pressure and light sensors.
The software package for the MSR175 includes 3 programs: MSR175 Dashboard, MSR ReportGenerator and MSR ShockViewer. The MSR dashboard allows you to configure the logger with just a few mouse clicks. The data recorded can be quickly transferred to a computer via the USB port. With a single click on the MSR ReportGenerator you can create a compact report. For detailed measured data analysis, you can use the MSR ShockViewer, which is integrated into the MSR175 software package as a free basic version.
Technical Data of the Sensors
| Internal Sensor | Working Range | Accuracy (Max Deviation) | Measurements / Storage Rate |
| 3-Axis Acceleration | ±15g and ±200g (-20~+65°C) | ±15g Sensor: ±0.15g (0~5g, +25°C), ±0.3 g (5~15g, +25°C) ±200g Sensor: ±2g (0~15g, +25°C), ±5g (15~100g, +25°C), ±10g (100~200g, +25°C) |
1600/s (±15%), 3200/s (±15%), 6400/s (±15%) |
| Temperature | -20~+65°C | ±0.5°C (-10~+65°C) | Every 10min |
| Relative Humidity | 0~100% rel. humidity (-20~+65°C) | ±2% rel. humidity (10~85%, 0~+40°C), ±4% rel. humidity (85~95%, 0~+40°C) | Every 10min |
| Air Pressure | 0~2000 mbar absolute (-20~+65°C) | ±2.5 mbar (750…1100 mbar absolute, +25°C) | Every 10min |
| Light | 0~83,000lx | Max sensitivity at 550nm | Every 10min |
Key Features: